Solution A of ICP-MS-MCS. 100 mL. Contains: (µg/mL) Antimony, Beryllium, Chromium, Manganese, Nickel, Titanium, Vanadium, Zinc at 20, Aluminium, Magnesium, Sodium, Sulfur/Sulphur at 1,000, Carbon at 2,000, Chlorine at 7,200 in 2% HNO3 + Trace HF. 12 months expiry date. Traceable to NIST 31XX series. ISO 9001:2015 certified, ISO/IEC 17025:2017 and ISO 17034:20166 accredited. Method 6020 CLP-M requires performance of a memory test after calibration of ICP-MS and before performing any analysis. This standard supports that test in the detection of the presence of any memory problems that would affect the data quality. To obtain the listed concentrations, equal volumes of Solutions A and B must be combined.
These cookies can't be rejected as they're required for essential functions of the website such as secure log-in or saving your progress during the ordering process.
Statistics
GoogleAnalytics
Google Analytics enables us to analyse website usage, helping us to change and optimise the site as necessary. All user data are anonymized.
Marketing
Google Ads
We use Google Ads to increase awareness for our company and to market our products. There is no advertising or marketing content from third parties on this website.